Mycardium AI Limited. Cleared by FDA on 30 June 2025 under submission K250670, product code QIH (automated radiological image processing software). What FDA's record says, in FDA's words.
Device record · K250670
- Device. EchoConfidence (USA)DEVICE
- Company. Mycardium AI Limited (GB)COMPANY
- Decision. 30 June 2025, 510(k), substantially equivalent (abbreviated review); received 5 March 2025DECISION
- Panel. RadiologyPANEL
- Product code. QIH, Automated Radiological Image Processing SoftwareCODE
- Regulation. 21 CFR 892.2050, Class IIREG
- Source. FDA database entry · decision document (7 pages)SOURCE
Retrieved · the fields above are FDA's; the statement below was copied from the decision document on that day.
What FDA cleared it for
The Indications for Use statement, verbatim, from the decision document.
EchoConfidence is Software as a Medical Device (SaMD) that displays images from a Transthoracic Echocardiogram, and assists the user in reviewing the images, making measurements and writing a report. The intended medical indication is for patients requiring review or analysis of their echocardiographic images acquired for their cardiac anatomy, structure and function. This includes automatic view classification; segmentation of cardiac structures including the left and right ventricle, chamber walls, left and right atria and great vessels; measures of cardiac function; and Doppler assessments. The intended patient population is both healthy individuals and patients in whom an underlying cardiac disease is known or suspected; the intended patient age range is for adults (>= 22 years old) and adolescent in the age range 18 – 21 years old. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
The product code's definition, from FDA's classification record. To provide automated radiological image processing and analysis tools. Software implementing artificial intelligence including nonadaptive machine learning algorithms trained with clinical and/or artificial data. In these devices, the algorithm training data typically impacts device performance. Adaptive AI algorithms are not within scope of this product code.
What the statement says
Pattern matches on the Indications for Use text. Each line is a fact that the text contains the words, not a finding about the device.
- Modalities namedultrasound
- Mentionsmeasurement or quantification
- Mentionsprescription use
- Mentionsadult patients
Adverse-event reports in MAUDE, by product code
As of , MAUDE held 10 reports filed against devices carrying product code QIH. That code is shared by every device of this type, not this device alone, and MAUDE reports are submitted by manufacturers, facilities and the public without FDA verification. The latest reports, in the reporter’s words:
2 July 2026 · Malfunction · SYNAPSE PACS · FUJIFILM HEALTHCARE AMERICAS CORPORATION · report 3004972322-2026-00002
ON 06/04/2026, FUJIFILM HEALTHCARE AMERICAS CORPORATION BECAME AWARE OF A BUG IN SYNAPSE PACS V7.4.310 THAT COULD POTENTIALLY LEAD TO A MISDIAGNOSIS IN THE CARDIOLOGY BASED STUDIES. IT WAS FOUND DURING REGRESSION TESTING OF THE PRODUCT THAT THE ANNOTATION TAKEN AND SAVED IN AN EARLIER VERSION OF THE PRODUCT CAN SHOW THE INCORRECT VALUE IN BOTH ZERO VIEWER (ZV) AND VX WHEN REOPENING THE STUDY IN THE NEWER VERSION. THIS ISSUE WAS INTRODUCED INTO THE PRODUCT DURING AN ENHANCEMENT TO SUPPORT […]
9 April 2026 · Malfunction · ORIGIN DATA MANAGEMENT (VERSION 3.2, SOFTWARE COMPONENT) · BRAINLAB SE · report 8043933-2026-00021
A PATIENT RECORD REQUIRED FOR STEREOTACTIC PLANNING COULD NOT BE FOUND DURING A FRAME-BASED STEREOTACTIC BIOPSY FOR A SUSPECTED BRAIN TUMOR. AFTER BRIEF TROUBLESHOOTING, IT BECAME CLEAR THAT THE PATIENT RECORD WAS UNINTENTIONALLY MERGED WITH ANOTHER PATIENT RECORD. THE SURGEON RESOLVED THE MERGE WITH SUPPORT FROM BRAINLAB. FURTHER INVESTIGATION BY BRAINLAB UNCOVERED FOUR MORE CASES OF INCORRECT PATIENT RECORD MERGES AT THE SAME SITE THAT WERE THEREAFTER RESOLVED. ACCORDING TO THE SURGEON […]
26 February 2026 · Malfunction · ORIGIN DATA MANAGEMENT (VERSION 3.2, SOFTWARE COMPONENT) · BRAINLAB SE · report 8043933-2026-00011
WHILE SEARCHING FOR A MISSING PATIENT RECORD IN THE BRAINLAB PATIENT SELECTION 6.1 APPLICATION, A CUSTOMER DISCOVERED THAT THE MISSING PATIENT RECORD HAD BEEN MERGED WITH AN UNRELATED PATIENT RECORD. THE CUSTOMER RESTORED THE MERGED PATIENT RECORD TO ITS ORIGINAL CONSTITUENT PATIENT RECORDS ON THE SAME DAY. AFTER THIS, THE MISSING PATIENT RECORD BECAME AVAILABLE IN THE PATIENT LIST. ACCORDING TO THE HOSPITAL, NO PATIENT WAS AFFECTED DUE TO THE DESCRIBED ISSUE AS NEITHER OF THE PATIENTS HAS HAD […]
13 August 2025 · Malfunction · ORIGIN DATA MANAGEMENT (VERSION 3.2, SOFTWARE COMPONENT) · BRAINLAB AG · report 8043933-2025-00058
A CUSTOMER REPORTED THAT A PATIENT RECORD DID NOT APPEAR IN THE BRAINLAB PATIENT SELECTION APPLICATION WHEN THE PATIENT'S NAME WAS SEARCHED. THE CUSTOMER FOUND THE MISSING PATIENT DATA MERGED INTO A DIFFERENT PATIENT RECORD AND WAS ABLE TO RESTORE THE PATIENT RECORDS TO THEIR ORIGINAL STATES. NO PATIENT WAS AFFECTED DUE TO THIS ISSUE.
25 June 2025 · Malfunction · SONIO DETECT · SONIO · report MW5171987
SONIO DETECT SOFTWARE AI ALGORITHM IS FAULTY AND WRONGLY LABELS FETAL STRUCTURES AND ASSOCIATES THEM WITH THE WRONG BODY PARTS. THE SYSTEM WILL FREQUENTLY DISPLAY AS VISUALIZED EVEN THOUGH IT HASN'T LEAVING PROVIDERS TO GIVE FALSE NORMALCY AND OPENING THEM UP TO LIABILITIES.
Where this fits
This page is a record, not medical advice and not a safety rating. It copies FDA's fields and FDA's statement, and lists what the statement's text mentions. Read the labeling before relying on any of it.
More devices under product code QIH, in Radiology, from Mycardium AI Limited, or cleared in 2025. The whole record is at FDA AI-enabled medical devices.