Radiology · 510(k) · 2021

OPTIS Mobile Next Imaging System, OPTIS Integrated Next Imaging System

Abbott Medical. Cleared by FDA on 8 June 2021 under submission K210458, product code NQQ (system, imaging, optical coherence tomography (oct)). What FDA's record says, in FDA's words.

Device record · K210458

  1. Device. OPTIS Mobile Next Imaging System, OPTIS Integrated Next Imaging SystemDEVICE
  2. Company. Abbott MedicalCOMPANY
  3. Decision. 8 June 2021, 510(k), substantially equivalent (traditional review); received 16 February 2021DECISION
  4. Panel. RadiologyPANEL
  5. Product code. NQQ, System, Imaging, Optical Coherence Tomography (Oct)CODE
  6. Regulation. 21 CFR 892.1560, Class IIREG
  7. Source. FDA database entry · decision document (10 pages)SOURCE

Retrieved · the fields above are FDA's; the statement below was copied from the decision document on that day.

What FDA cleared it for

The Indications for Use statement, verbatim, from the decision document.

The Ultreon 1.0 Software is intended to be used only with compatible OPTIS Next Imaging Systems. The OPTIS Next Imaging System with a compatible Dragonfly OPTIS Imaging Catheter or Dragonfly OpStar Imaging Catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The Dragonfly OPTIS Imaging Catheter or Dragonfly OpStar Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The Dragonfly OPTIS Imaging Catheter or Dragonfly OpStar Imaging Catheter is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure. The OPTIS Next Imaging System is intended for use in the catheterization and related cardiovascular specialty laboratories and will further compute and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices. The physician may use the acquired physiological parameters, along with knowledge of patient history, medical expertise, and clinical judgment to determine if therapeutic intervention is indicated. The OPTIS Mobile Next and OPTIS Integrated Next with a compatible Dragonfly OPTIS or Dragonfly OpStar Imaging Catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The Dragonfly OPTIS or Dragonfly OpStar Imaging Catheter is […]

K210458, FDA decision document · source

The product code's definition, from FDA's classification record. This device uses back-reflected light to create two-dimensional images versus back-reflected sound waves as are used in ultrasound imaging.

What the statement says

Pattern matches on the Indications for Use text. Each line is a fact that the text contains the words, not a finding about the device.

Adverse-event reports in MAUDE, by product code

As of , MAUDE held 69 reports filed against devices carrying product code NQQ. That code is shared by every device of this type, not this device alone, and MAUDE reports are submitted by manufacturers, facilities and the public without FDA verification. The latest reports, in the reporter’s words:

12 June 2026 · Malfunction · G10-01 GENTUITY HF-OCT IMAGING SYSTEM · GENTUITY, LLC · report 3011578544-2026-00054

ON14 FEB 2023: (B)(6) HEAD OFFICE: ON (B)(6) 2023 AFTER UPGRADING S/N: (B)(6) (V21.11.7 V21.11.9) AND THE BIOS UPDATE WORK, "CRITICALERROR" OCCURRED WHEN STARTING THE CONSOLE AND LOGGING INTO THE SERVICE BACKDOOR. UPON REBOOTING AFTER SHUTDOWN, THE MESSAGE DISPLAYED "THERE WAS AN ERROR STARTING THE SYSTEM. PLEASE CONTACT GENTUITY IF THE PROBLEM PERSISTS." AFTER THAT, REPEATED MANY TIMES RESTARTS, BUT SAME MESSAGE DISPLAYED ALL. "THERE WAS AN ERROR STARTING THE SYSTEM. PLEASE CONTACT GENTUITY […]

12 June 2026 · Malfunction · G30-01 VIS-RX MICRO-IMAGING CATHETER · GENTUITY, LLC · report 3011578544-2026-00056

BOTH DOCTORS (B)(6) AT (B)(6) SAID THAT THEY ARE CONCERNED ABOUT THE ACCURACY OF THE LENGTH OF OUR SYSTEM. DR (B)(6) SAID THAT "THEY (THE GROUP) ARE ALL TALKING ABOUT IT, AND THEY DON'T BELIEVE IT." WHEN I DID A CASE WITH DR (B)(6) HE PLACED A 23 MM ABBOTT STENT, AND WHEN I MEASURED IT, THE SYSTEM MEASURED 27 MM. SO I WENT BACK AND LOOKED AT DOCTOR (B)(6) CASE FROM THE DAY PRIOR, AND THAT WAS A 26 MM BIOTRONIC STENT, AND IT MEASURED 26 MM EXACTLY. (I DID NOT GET THE OPPORTUNITY TO SHARE THIS […]

1 June 2026 · Malfunction · GENTUITY HF-OCT IMAGING SYSTEM · GENTUITY, LLC · report 3011578544-2026-00040

DR. (B)(6), PART A PA STUDY: THE FIRST ATTEMPT OF AN ACQUISITION TIME OUT ON CALIBRATION AND THE STAFF TRIED IT A SECOND TIME WITH A SUCCESSFUL IMAGE ACQUIRED. NO INTERVENTION WAS PERFORMED, SO THE CASE WAS ENDED. POST PROCEDURE, (B)(6), (RESEARCH COORDINATOR) ATTEMPTED TO RELEASE THE VIS RX CATHETER FROM THE PIM AND THE PULLBACK PORTION OF THE CATHETER WOULD NOT RELEASE. (B)(6) CALLED ME AGAIN ON WHATSAPP VIDEO CHAT AND I TALKED HER THROUGH TROUBLE SHOOTING WITH NO SUCCESS. I ASKED HER TO […]

28 May 2026 · Malfunction · GENTUITY HF-OCT IMAGING SYSTEM · GENTUITY, LLC · report 3011578544-2026-00031

A RECENT CONVERSATION AMONGST CLINICAL SPECIALISTS PROMPTED ME TO REVIEW EACH RUN IN THE CONSOLE FOR REPETITIVE/STACKED FRAMES AND I FOUND THAT IN 26 OF THE 28 RUNS FROM 16 CASES HERE HAD DISTAL FRAMES STACKED/FAILURE TO PULL BACK BEFORE GENERATING IMAGES.

27 May 2026 · Malfunction · GENTUITY HF-OCT IMAGING SYSTEM · GENTUITY, LLC · report 3011578544-2026-00027

WHILE DOING A CASE WITH DR (B)(6) YESTERDAY, HE WAS VERY UNHAPPY WITH THE LONGITUDINAL MEASUREMENTS. HE DID NOT BELIEVE THE PRE (WHICH I MEASURED AT 26 MM, AND HE THOUGHT IT REALLY SHOULD BE 28 OR EVEN 30MM). EVEN TO THE POINT THAT HE WANTED TO DISCUSS AFTER THE CASE. (IT WAS TOWARDS THE DISTAL END OF THE RUN). THE POST STENT MEASUREMENTS WERE OFF ALSO, WHICH I DID NOT BRING TO HIS ATTENTION. HE HAD PLACED A 28MM ABBOTT STENT, AND IT MEASURED 31MM. BOTH OCT RUNS AND LOGS WERE UPLOADED TO BOX.

Where this fits

This page is a record, not medical advice and not a safety rating. It copies FDA's fields and FDA's statement, and lists what the statement's text mentions. Read the labeling before relying on any of it.