Radiology · 510(k) · 2025

SCENARIA View Phase 5.0

FUJIFILM Corporation. Cleared by FDA on 20 May 2025 under submission K250370, product code JAK (system, x-ray, tomography, computed). What FDA's record says, in FDA's words.

Device record · K250370

  1. Device. SCENARIA View Phase 5.0DEVICE
  2. Company. FUJIFILM Corporation (JP)COMPANY
  3. Decision. 20 May 2025, 510(k), substantially equivalent (traditional review); received 10 February 2025DECISION
  4. Panel. RadiologyPANEL
  5. Product code. JAK, System, X-Ray, Tomography, ComputedCODE
  6. Regulation. 21 CFR 892.1750, Class IIREG
  7. Source. FDA database entry · decision document (11 pages)SOURCE

Retrieved · the fields above are FDA's; the statement below was copied from the decision document on that day.

What FDA cleared it for

The Indications for Use statement, verbatim, from the decision document.

The SCENARIA View system is indicated to acquire axial volumes of the whole body including the head. Images can be acquired in axial, helical, or dynamic modes. The SCENARIA View system can also be used for interventional needle guidance. Volume datasets acquired by a SCENARIA View system can be post-processed in the SCENARIA View system to provide additional information. Post-processing capabilities of the SCENARIA View software include multi-planar reconstruction (MPR), and volume rendering. Volume datasets acquired by a SCENARIA View system can be t ransferred to external devices via a DICOM standard interface. The Low Dose CT Lung Cancer Screening Option for the SCENARIA View system is indicated for using low dose CT for lung cancer screening. The screening must be conducted with the established program criteria and protocols that have been approved and published by a governmental body, a professional medical society, and/or FUJIFILM Corporation. The SCENARIA View system is intended for general populations. Indications for Use Comparison No difference in indications for use. P a g e | 3 of 7

K250370, FDA decision document · source

What the statement says

Pattern matches on the Indications for Use text. Each line is a fact that the text contains the words, not a finding about the device.

Adverse-event reports in MAUDE, by product code

As of , MAUDE held 5,281 reports filed against devices carrying product code JAK. That code is shared by every device of this type, not this device alone, and MAUDE reports are submitted by manufacturers, facilities and the public without FDA verification. The latest reports, in the reporter’s words:

25 September 2026 · Malfunction · REVOLUTION EVO · GE MEDICAL SYSTEMS, LLC · report 50896044

The GE CT table wouldn't move during stroke. Patient had to be physically moved to resume scan, delay in care. GE CT table has had the same issues with the table working intermittently x 3 weeks. Multiple techs have worked on it and the same problem persists.

25 September 2026 · Malfunction · Needle Guide · Quantum Surgical SAS · report 3023140882-2026-00002

23 September 2026 · Malfunction · ILLUMISITE · COVIDIEN LP - SUPERDIMENSION INC · report 3004962788-2026-00087

It was reported that during the procedure, one locatable guide indicated as outside of the sensor sensing volume and another locatable guide had difficulty retracting from the extended working channel. The physician completed the superDimension portion of the case. No patient complications have been reported as a result of this event.

23 September 2026 · Other · NL4000 · Samsung HME America, Inc. · report 3004938766-2026-00043

23 September 2026 · Malfunction · ILLUMISITE · COVIDIEN LP - SUPERDIMENSION INC · report 3004962788-2026-00088

Where this fits

This page is a record, not medical advice and not a safety rating. It copies FDA's fields and FDA's statement, and lists what the statement's text mentions. Read the labeling before relying on any of it.